Accuracy of IOL Power Calculation Formulas for Quadrifocal Acrysof IQ PanOptix TFNT Implantation in Patients With Previous Corneal Refractive Surgery: Comparison of SS-OCT–Based Biometers
카이안과 | 2022-01-24
Accuracy of IOL Power Calculation Formulas
for Quadrifocal Acrysof IQ PanOptix TFNT Implantation 
in Patients With Previous Corneal Refractive Surgery
:Comparison of SS-OCT–Based Biometers
 
Hyunggoo Kwon, MD, Sohee Jeon, MD, PhD, Ayoung Choi, BSc, Optom,
 
 
이전 각막 굴절 수술을 받았던 환자에게
4중 초점 인공수정체 Acrysof IQ Panoptix TFNT IOL 삽입을 위한
도수 계산 공식의 정확도 : SS-OCT 기반 생체측정기 비교
 
[카이안과 의료팀] 권형구 원장, 전소희 원장, 최아영 검안사
 
 
 
 

 
 

과거, 각막 굴절 수술을 받았던 
시각적으로 특별한 의미가 있는 백내장 환자에게 
4중 초점 인공수정체 Acrysof IQ Panoptix TFNT IOL을 삽입하기 위해 
광간섭단층촬영장비(Swept-Source OCT)를 사용하여 
두 생체 측정기의 인공 수정체(IOL) 도수 계산 공식의 정확도를 평가한 연구
로,
 
각막 굴절 수술 및 Acrysof IQ Panoptix TFNT 인공 수정체 삽입의 병력이 있는 
50명의 환자 데이터를 기반으로 평가한 자료에 따르면, 

이전에 각막굴절 수술을 받은 환자에서 
TFNT 인공수정체 삽입술의 시력 및 굴절 예후는 양호하였으며
 IOLMaster 700의 Barrett True-K 공식은
 TFNT 인공수정체 삽입술에서 최고의 굴절 결과를 보여주었습니다.
 
 

 
 

PURPOSE:

To evaluate the accuracy of intraocular lens (IOL) power calculation formulas

from two biometers using swept-source optical coherence tomography

for quadrifocal Acrysof IQ Panoptix TFNT IOL (Alcon Laboratories,

Inc) implantation in patients with visually significant cataract with previous corneal refractive surgery.

 

METHODS:

This retrospective study comprised 50 eyes from 50 patients

with a history of corneal refractive surgery and TFNT IOL implantation.

Candidate formulas were Shammas-PL and Barrett True-K in the Argos (Movu, Inc),

Barret True-K and Haigis-L in the IOLMaster 700 (Carl Zeiss Meditec AG),

and Haigis using the total keratometry (TK) mode in the IOLMaster 700.

The main outcome measure was the mean absolute error (MAE) detected at postoperative 6 months.

The refractive accuracy was also evaluated as number and percentage of eyes within ±0.25, ±0.50, and ±0.75 diopters (D) of the prediction error.

 

RESULTS:

The uncorrected distance and near visual acuity were 0.32 ± 0.34 and 0.46 ± 0.20 logMAR at baseline,

and significantly improved to 0.04 ± 0.07 and 0.03 ± 0.06 logMAR at postoperative 6 months (P < .001 for all analysis)

with a mean spherical equivalent of −0.20 ± 0.39 D. The MAE was smallest for the Barrett True-K formula in the IOLMaster 700 (0.36 ± 0.26 D)

and largest for the Shammas-PL formula in the Argos (0.59 ± 0.37 D). The Barrett True-K formula from both devices showed that 90% of eyes were within ±0.75 D of MAE.

 

CONCLUSIONS:

The visual and refractive outcomes of TFNT IOL implantation in patients

with previous corneal refractive surgery were favorable. The Barrett True-K formula

in the IOLMaster 700 showed the best refractive outcome for TFNT IOL implantation.

 
 
 
약도전송 페이스북 네이버 블로그 인스타그램 유튜브 카카오플러스

개인정보 취급방침

1.개인정보 수집이용에 대한 동의 ‘카이안과’는 회원님의 개인정보를 중요시하며, “정보통신망 이용촉진 및 정보보호”에 관한 법률을 준수하고 있습니다. 회원님의 소중한 개인정보는 다음과 같은 정책에 따라 수집 및 이용되며, 회원님께서 제공하시는 개인정보가 어떠한 용도와 방식으로 이용되고 있으며, 개인정보보호를 위해 어떠한 조치가 취해지고 있는지 알려드립니다. 또한 개인정보취급방침을 개정하는 경우 웹사이트 공지사항을 통하여 공지할 것입니다.


2.개인정보의 수집•이용 목적 회원님께서 요청하신 제휴 문의에 대한 질의에 더욱 정확한 답변을 위해 성함, 연락처, 상담내용 등의 정보가 수집됩니다. - 수집항목 : 성명, 연락처, 상담내용 - 이용목적 : 상담내용에 대한 답변, 상담 관련 마케팅 및 정보안내 고지에 활용


3.보유이용 기간 원칙적으로, 개인정보 수집 및 이용목적이 달성된 후에는 해당 정보를 바로 파기합니다. 단, 다음의 정보에 대해서는 아래의 사유로 명시한기간 동안 보존합니다. - 보존 이유 : 회원님의 동의를 통한 정보 유지 - 보존 기간 : 회원정보 삭제 요청 시까지

오시는길 문자로 받기
카카오톡 위치전송